Camilla L Duus, MB
Bispebjerg Hospital and Rigshospitalet
Wearable sensors recorded heart rate, breathing rate and blood oxygen saturation for up to four days after surgery. Researchers compared these recordings with routine Early Warning Score measurements. Sensor alarms were disabled.
Continuous monitoring detected oxygen saturation below 92% in 98% of patients, compared with 16% using routine measurements. It also detected more episodes of rapid heart rate. This pilot assessed detection, rather than whether monitoring prevents complications.