WARD Medical Validation

WARD Medical Validation

Principal Investigator

Christian S. Meyhoff, MD, PhD, Professor

Study design
Prospective study/cohort
Sample size
20 participants
Study sites

Bispebjerg Hospital

Inclusion criteria
  • Adult patients admitted with AECOPD
Study procedures

Researchers compared wireless sensor readings with standard hospital monitors during two hours of simultaneous monitoring, including observed breathing counts.

Results

180 sample-pairs of vital signs from 20 with AECOPD patients were recorded for comparison. The wireless versus standard monitor bias was 0.03 (LoA −3.2 to 3.3) bpm for HR measurements, 1.4% (LoA −0.7% to 3.6%) for SpO2, −7.8 (LoA −22.3 to 6.8) mmHg for systolic BP and −6.2 (LoA −16.8 to 4.5) mmHg for diastolic BP. The wireless versus standard monitor bias for RR measurements was 0.75 (LoA −6.1 to 7.6) brpm.

Outcome

Heart rate measurements agreed well with standard monitoring. Agreement for blood oxygen saturation was borderline acceptable. Breathing rate and blood pressure measurements required more cautious interpretation.

Trial registration
NCT04248842